Services Available |
---|
107-LC 的热敏电阻置于环氧基树脂填充的铝外壳中。 测量范围为-35° 至 +50°C,-5° 至 +95°C测量范围请参考108-LC 。
The 107-LC consists of a thermistor encapsulated in an epoxy-filled aluminum housing. Its cable terminates in a connector that attaches to the ET107 Weather Station, ET106 Weather Station (retired), or MetData1 Weather Station (retired). The 107-LC measures the range of -35° to +50°C; to measure the -5° to +95°C temperature range, see the 108-LC probe.
The 107-LC is suitable for shallow burial only. It should be placed horizontally at the desired measurement depth to avoid thermal conduction from the surface to the thermistor. Placement of the sensor cable inside a rugged conduit may be advisable for long cable runs, especially in locations subject to digging/mowing, traffic, use of power tools, or lightning strikes.
p>The 107-LC can be submerged to 50 feet. Please note that the probe is not weighted. Therefore, the installer should either add a weighting system or secure the probe to a fixed, submerged object, such as a piling.
The cable of the 107-LC provides a connector that attaches to the ET107 Weather Station, ET106 Weather Station (retired), or MetData1 Weather Station (retired). The ET107 Weather Station has an on-board CR1000M module. Both the MetData1 Weather Station and ET106 Weather Station have an on-board CR10X module.
107-LC: 7
展开全部收起全部
The thermistor is located approximately 3 mm (0.125 in.) back from the probe tip.
When these sensors are purchased, the following calibration services are offered: TEMPCAL and TEMPCAL2.
For both of these services, calibration can be made at different values if it is requested by the purchaser at the time of purchase. In addition, both of these calibration services can be requested after sensor purchase using a return material authorization (RMA) number. To request an RMA number, refer to the Repair and Calibration page.
The sensor/probe consists of a non-linear thermistor configured with a precision resistor in a half-bridge circuit, as shown in the product’s manual:
To measure the sensor/probe, the measurement device has to provide a precision excitation voltage (Campbell Scientific data loggers use 2000 mV), measure the voltage across the precision resistor, determine the thermistor resistance (Ohm's law), and convert the resistance to temperature using the Steinhart-Hart equation.
The Steinhart-Hart equation is 1/T = A + Bln(R) + C(ln(R))3 where:
For the 107-L, 107-LC, 108-L, and 108-LC, the following are the coefficients for the Steinhart-Hart equation:
For the 109-L, the following are the coefficients for the Steinhart-Hart equation:
Note the difference between calibration and a field check. Calibration cannot be done in the field, as it requires an experienced technician and specialized equipment.
Field checks of measurements can be done to determine if the data make sense with the real-world conditions. Follow these steps to field check a sensor:
大多数 Campbell Scientific 传感器的型号中都带有 –L,它表示用户指定线缆长度。如果传感器型号名称列有 –LX (这里 “X” 是其它的某个字符), 那么该传感器需要用户指定长度,但线缆尾端会配有用于某个独特系统的特殊快速接头:
如果一个传感器的主型号数字的后面没有被指定 –L 或其它的 –LX 字符,那么该传感器将具有固定的线缆长度。在产品页面订购栏中 (Ordering tab) 的描述字段的末尾,会列出线缆的长度。例如 034B-ET 型号的描述字段含有信息:Met One风传感器适用于ET气象站,67英寸线缆 (Met One Wind Set for ET Station, 67 inch Cable)。产品按固定的线缆长度做尾端处理,默认是尾线方式。
如果线缆尾端配有用于某个独特系统的特殊快速接头,则型号尾部的字符会指定该传感器用于哪一个系统。例如,034B-ET 型号表明该传感器是一个 034B 且用于 ET107 系统。
不是每一种传感器都有不同的线缆尾端选项。通过查找传感器产品页面的订购栏 (Ordering tab) 中的两个位置,可以检查某种特定传感器的可用的线缆尾端选项:
如果传感器以 –ET, –ETM, –LC, –LQ, 或 –QD 等版本的型号供应,那么线缆尾端选项已经反映在该传感器的产品型号中。例如,034B 以 034B-ET, 034B-ETM, 034B-LC, 034B-LQ, 和 034B-QD 等型号供应。
所有其它的线缆尾端选项,如果可用,会列在该传感器产品页面的订购栏 (Ordering tab) 中的线缆尾端选项 (“Cable Termination Options) 区域。例如,034B-L 风速风向传感器具有 –CWS, –PT, 和 –PW 等线缆尾端选项,显示在 034B-L 产品页面的 订购栏 (Ordering tab) 位置。
注: 当更新的产品添加到我们的库存中时,一般来说,我们会在单个传感器的产品型号下面列出多种线缆尾端选项,而不是创建多个产品型号。例如,HC2S3-L 具有 –C 线缆尾端选项用于连接到 CS110,而我们并没有使用 HC2S3-LC 产品型号。
Campbell Scientific 的很多传感器都有可行的不同线缆尾端处理选项,包括以下几种:
注: 线缆尾端选项的可行性因传感器而异。例如,传感器可能会有 无选项、两个选项或几个选项可供选择。如果预期的选项没有列在所指定的传感器中,请联系 Campbell Scientific 的应用工程师寻求帮助。