Campbell Scientific的CS525 pH值探头提供了可靠的精确的pH值测量,在整个工作温度范围内都实施补偿。该探头通常测量水媒的或半固态溶液的pH值,探头可以被浸入或插入水箱、水管或开放水渠中。
CS525采用SENTRON的高科技的离子敏感场效应晶体管(ISFET)半导体作为它的pH感应元件,并且包含 银/氯化银-氯化钾参考系统。ISFET技术是当今最强大的pH监测技术,它相当程度地减小了极端pH条件下的酸性或碱性的数值误差。CS525可以监测含有高固体组分、腐蚀性化学品、或者容易阻塞或污染传统的玻璃泡pH计的生物体的液体的pH值。
CS525的耐用设计,使得它适合任何液体的pH值监测应用,从实验室到严酷的野外应用。探头的电子单元安全地封装在耐用的PEEK外壳里。不需要玻璃泡,也消除了玻璃破碎的可能性,使得CS525更耐用更安全。
Note: CS525-L ISFET pH值传感器在接地环路和流电条件下高度易受影响。克服该问题的最好隔离技术是使用CWS900系列无线传感器接口。在很多情况下,CS525也可以实现隔离,通过不接地到数据采集器,从而连不到大地。请在订购传感器之前与Campbell Scientific应用工程师讨论您的应用。 |
CE Compliance | CE Compliant |
pH Range | 0 to 14 |
Power Requirements | 4.5 to 5.5 Vdc |
Current Consumption | 2 mA (nominal) |
Measurement Time | 200 ms < 67% < 500 ms |
Accuracy | ±0.1 pH (with two-point calibration) |
Operating Temperature | 0° to 70°C |
Allowed Water Pressure | 0 to 30 kPa (0 to 45 psi) |
Cable Type | Two-twisted pair, 22 AWG cable with Santoprene jacket |
Sensor Material | Polyaryletheretherketone (PEEK) |
Diameter | 16 mm (0.63 in.) |
Maximum Cable Length | 300 m (1000 ft) |
Length | 190 mm (7.48 in.) |
Weight | 318 g (11.2 oz) with 3.05-m (10-ft) cable |
Output |
|
-NOTE- | Output range used depends on which wires are connected to the data logger. |
Range A |
|
Range B | ±400 mV with a sensitivity of -45 mV to -50 mV/pH at 25°C |
Please note: The following shows notable compatibility information. It is not a comprehensive list of all compatible products.
Product | Compatible | Note |
---|---|---|
CR200X (retired) | The CR200X-series dataloggers require the VDIV2:1 2-to-1 Voltage Divider; contact Campbell Scientific about using this sensor with a retired data logger. | |
CR206X (retired) | The CR200X-series dataloggers require the VDIV2:1 2-to-1 Voltage Divider; contact Campbell Scientific about using this sensor with a retired data logger. | |
CR211X (retired) | The CR200X-series dataloggers require the VDIV2:1 2-to-1 Voltage Divider; contact Campbell Scientific about using this sensor with a retired data logger. | |
CR216X (retired) | The CR200X-series dataloggers require the VDIV2:1 2-to-1 Voltage Divider; contact Campbell Scientific about using this sensor with a retired data logger. | |
CR295X (retired) | The CR200X-series dataloggers require the VDIV2:1 2-to-1 Voltage Divider; contact Campbell Scientific about using this sensor with a retired data logger. | |
CR3000 (retired) | ||
CR5000 (retired) | ||
CR800 (retired) | ||
CR850 (retired) |