034B将3-杯风速计和风向标组成一个单独的组件来测量风速和风向。其线缆可与我们的数据采集器连接,可应用于多种应用的测量。
034B 利用带有密封的磁簧开关的3-杯风速计测量风速。风杯的旋转产生与风速成正比的脉冲。数据采集器脉冲计数器可测量出脉冲的频率,然后将其转换成工程单位,例如mph, m/s, knots。
风向由电位计检测。随着精密的激发电压从数据记录器施加到电位计元件,输出信号是模拟电压,其与风向方位角成正比。
Applications | General (Rain with light snow. Little or now riming or blowing sand. No salt spray.) |
Sensor | 3-cup anemometer and vane |
Measurement Description | Wind speed and direction |
Operating Temperature Range | -30° to +70°C |
Weight | 907 g (2.0 lb) |
Wind Speed (Anemometer) |
|
Range | 0 to 75 m/s (0 to 167 mph) |
Starting Threshold | 0.4 m/s (0.9 mph) |
Sensor Output | Pulsed contact closure |
Accuracy |
|
Resolution | (0.7998 m s–1) / (scan rate in seconds) or (1.789 mph) / (scan rate in seconds) |
Anemometer Radius | 10.7 cm (4.2 in.) |
Anemometer Height | 24.4 cm (9.6 in.) |
Output Signal | Contact closure (reed switch) |
Wind Direction (Vane) |
|
Mechanical Range | 360° |
Electrical Range | 356° (4° open) |
Accuracy | ±4° |
Damping Ratio | 0.25 |
Resolution | < 0.5° |
Potentiometer Resistance | 0 to 10 kΩ (open at crossover) |
Threshold | 0.4 m s–1 (0.9 mph) |
Vane Length | 33.5 cm (13.2 in.) |
The 034B can be attached to a Campbell Scientific crossarm using a 17953 Nu-Rail fitting or a CM220 Right Angle Mounting Bracket. Alternatively, the 034B can be attached to the top of our stainless-steel tripods via the CM216 Sensor Mounting Kit. The CM216 extends 4 inches above the mast of a CM106B, CM110, CM115, or CM120 tripod.
The 034B's anemometer uses one pulse count channel on the data logger. Its wind vane requires one single-ended channel and access to an excitation channel (the excitation channel can be shared with other high impedance sensors).
The 034B's anemometer is measured by the PulseCount Instruction in CRBasic, and by Instruction 3 (Pulse Count) in Edlog. The wind vane is measured by the BrHalf Instruction in CRBasic and by Instruction 4 (Excite-Delay-SE) in Edlog. The measurements are typically processed for output with the Wind Vector instruction (not present in the CR500 or CR9000).
034B-L: 14
展开全部收起全部
Wind vane tails are user-replaceable and are available for purchase as pn 13765, Replacement Wind Vane Tail for 034B or 034A.
No. A data logger is needed to measure and store data from the 034B-L.
The supply voltage is 5 Vdc.
They are completely interchangeable, and no changes are needed.
The sensor does not consume power to operate.
For proper siting of the wind sensor, refer to the 034B-L manual. The 034B-L can be mounted using a 1.00 IPS pipe of any length.
The 034B-L uses a switch closure for wind speed measurement, and the 03002-L uses a low-level ac output for wind speed measurement. Both use a pulse channel on the data logger to measure wind speed.
The information included on a calibration sheet differs with each sensor. For some sensors, the sheet contains coefficients necessary to program a data logger. For other sensors, the calibration sheet is a pass/fail report.
This depends on the information contained in the calibration sheet: