HFP01SC 测量土壤热通量,通常用于能量平衡系统或波文比通量系统。倾向用于那些要求最高可能测量精度的应用。每个站点至少需要2个土壤热通量传感器,才能提供足够的空间平均。带有非均匀介质的站点可能需要更多的传感器。
HFP01SC 包含一个热电堆和一个薄膜加热器。热电堆来测量通过传感器板上的温度梯度。在原位野外校准期间,加热器生成通过传感器板的热通量,其功率大小由数据采集器测量;数据采集器同时测量传感器的响应信号,再计算生成新的校准常数。出厂时,每一个热通量板传感器都会单独校准,输出热通量。
自校准修正了由传感器与周围介质之间热传导率的差异、温度变化以及轻微的传感器不稳定性等因素所导致的误差。
Sensor Type | Thermopile with film heater |
Sensitivity | 50 μV W-1 m-2 (nominal) |
Nominal Resistance | 2 Ω |
Temperature Range | -30° to +70°C |
Expected Typical Accuracy | ±3% of reading |
Heater Resistance | 100 Ω (nominal) |
Heater Voltage Input | 9 to 15 Vdc |
Heater Voltage Output | 0 to 2 Vdc |
Duration of Calibration | ±3 minutes @ 1.5 W (typically performed every 3 to 6 hours) |
Average Power Consumption | 0.02 to 0.04 W |
Plate Diameter | 80 mm (3.15 in.) |
Plate Thickness | 5 mm (0.20 in.) |
Weight | 200 g (7.05 oz) without cable |
CR1000X program that measures the HFP01SC, performs the self-calibration, and checks for calibration validity. Refer to the manual for the equations used for the self-calibration and calibration-validity checks. A table in the manual provides a cross reference of the terms in the equations in the manual with the constants and variables in the example data logger program.
HFP01SC-L: 13
展开全部收起全部
No. The HFP01SC-L must be in full contact with the media. Railroad ballast is too coarse.
Rather than using a running average to find the millivolt output during a calibration, use a single sample with 50 or 60 Hz integration. See Example 1 in the 2014 or later version of the HFP01SC-L manual.
The example CRBasic program runs in either SequentialMode or PipeLineMode. To force the CRBasic program to run in PipeLineMode, add the instruction PipeLineMode to the beginning of the program.
A calibration shift occurs if the HFP01SC-L is not making full contact with the soil during the calibration cycle. The following could cause the plate to lose contact with the soil: a soil freeze/thaw cycle, soil swelling/contracting because of extreme drying/wetting cycles, or rodents burrowing past the plate.
The in-situ calibration is helpful for quality assurance/quality control. The multiplier determined from the in-situ calibration should be within ±10% of the factory-determined calibration. If it is not, the plate may be damaged, not wired correctly to the data logger, or not making full contact with the soil.
The information included on a calibration sheet differs with each sensor. For some sensors, the sheet contains coefficients necessary to program a data logger. For other sensors, the calibration sheet is a pass/fail report.
This depends on the information contained in the calibration sheet:
大多数 Campbell Scientific 传感器的型号中都带有 –L,它表示用户指定线缆长度。如果传感器型号名称列有 –LX (这里 “X” 是其它的某个字符), 那么该传感器需要用户指定长度,但线缆尾端会配有用于某个独特系统的特殊快速接头:
如果一个传感器的主型号数字的后面没有被指定 –L 或其它的 –LX 字符,那么该传感器将具有固定的线缆长度。在产品页面订购栏中 (Ordering tab) 的描述字段的末尾,会列出线缆的长度。例如 034B-ET 型号的描述字段含有信息:Met One风传感器适用于ET气象站,67英寸线缆 (Met One Wind Set for ET Station, 67 inch Cable)。产品按固定的线缆长度做尾端处理,默认是尾线方式。
如果线缆尾端配有用于某个独特系统的特殊快速接头,则型号尾部的字符会指定该传感器用于哪一个系统。例如,034B-ET 型号表明该传感器是一个 034B 且用于 ET107 系统。
不是每一种传感器都有不同的线缆尾端选项。通过查找传感器产品页面的订购栏 (Ordering tab) 中的两个位置,可以检查某种特定传感器的可用的线缆尾端选项:
如果传感器以 –ET, –ETM, –LC, –LQ, 或 –QD 等版本的型号供应,那么线缆尾端选项已经反映在该传感器的产品型号中。例如,034B 以 034B-ET, 034B-ETM, 034B-LC, 034B-LQ, 和 034B-QD 等型号供应。
所有其它的线缆尾端选项,如果可用,会列在该传感器产品页面的订购栏 (Ordering tab) 中的线缆尾端选项 (“Cable Termination Options) 区域。例如,034B-L 风速风向传感器具有 –CWS, –PT, 和 –PW 等线缆尾端选项,显示在 034B-L 产品页面的 订购栏 (Ordering tab) 位置。
注: 当更新的产品添加到我们的库存中时,一般来说,我们会在单个传感器的产品型号下面列出多种线缆尾端选项,而不是创建多个产品型号。例如,HC2S3-L 具有 –C 线缆尾端选项用于连接到 CS110,而我们并没有使用 HC2S3-LC 产品型号。