Services Available | |
---|---|
Repair | No |
Calibration | Yes |
Free Support | Yes |
CSAT3三维超声风速仪可以测量正交的三维风速和声音速度。在涡度相关系统中,CSAT3用来测量水平和垂直风的湍流波动;它们继而用于计算动量通量和摩擦风速。CSAT3还可提供水平风速和风向的平均测量。
CSAT3的测量路径垂直方向长 10 cm,仪器以脉冲式的超声波模式工作。输出正交的三维风速分量(ux, uy, uz)和声速,最大测量速率为 60 Hz。提供模拟输出和两种数字输出接口。
CSAT3的测量可由 CSAT3’的内部时钟、PC生成的RS-232指令或者是数据采集器的 SDM指令来触发。SDM协议支持群组触发,可以同步多个CSAT3三维超声风速仪。
FW05微丝瞬时响应热电偶(12.7 μm 直径)可作为快速响应温度测量的可选传感器。
Measurement Path Length |
|
Path Angle from Horizontal | 60° |
Construction | Sealed sonic transducers and electronics |
Anemometer Head Materials | Stainless-steel tubing |
Electronics Box Materials | Cast aluminum |
Operating Temperature Range | -30° to +50°C |
Voltage Supply | 10 to 16 Vdc |
Current |
|
Digital SDM Output Signal | CSI 33.3 k baud serial interface for data logger/sensor communication. (Data type is 2-byte integer per output plus 2-byte diagnostic.) |
Support Arm Diameter | 1.59 cm (0.63 in.) |
Transducer Diameter | 0.64 cm (0.25 in.) |
Transducer Mounting Fingers Diameter | 0.84 cm (0.33 in.) |
Cable Length | 7.62 m (25 ft) |
Anemometer Head Dimensions | 47.3 x 42.4 cm (18.6 x 16.7 in.) |
Anemometer Head Weight | 1.7 kg (3.7 lb) |
Electronics Box Dimensions | 26 x 16 x 9 cm (10.24 x 6.3 x 3.54 in.) |
Electronics Box Weight | 3.8 kg (8.4 lb) |
Measurements |
|
Outputs |
ux, uy, uz, c (ux, uy, uz are wind components referenced to the anemometer axes; c is speed of sound.) |
Speed of Sound | Determined from three acoustic paths; corrected for crosswind effects. |
Measurement Rate | Programmable from 1 to 60 Hz, instantaneous measurements. Two over-sampled modes are block averaged to either 20 Hz or 10 Hz. |
Measurement Resolution |
|
Offset Error |
|
Gain Error |
|
Wind Direction Accuracy | ±0.7° at 1 m/s (for horizontal wind) |
Rain | Innovative ultrasonic signal processing and user-installable wicks considerably improve the performance of the anemometer under all rain events. |
Digital RS-232 Output Signal |
|
Baud Rate | 9600, 19200 bps |
Data Type | 2-byte integer per output plus 2-byte diagnostic |
Analog |
|
Number of Outputs | 4 |
Voltage Range | ±5 V |
Number of Bits | 12 |
SDM & RS-232 Digital Outputs Reporting Range |
|
Full-Scale Wind | ±65.535 m/s autoranging between four ranges (Least significant bit is 0.25 to 2 mm/s.) |
Speed of Sound | 300 to 366 m/s (-50° to +60°C) Least significant bit is 1 mm/s (0.002°C). |
Analog Outputs Reporting Range |
|
ux | ±30 m s-1, ±60 m s-1 |
uy | ±30 m s-1, ±60 m s-1 |
uz | ±8 m s-1 |
c | 300 to 366 m s-1 (-50 to +60°C) |
Analog Outputs LSB |
|
ux | 15 mm s-1, 30 mm s-1 |
uy | 15 mm s-1, 30 mm s-1 |
uz | ±8 m s-1 |
c | 16 mm s-1 (0.026°C) |
Please note: The following shows notable compatibility information. It is not a comprehensive list of all compatible products.
Product | Compatible | Note |
---|---|---|
21X (retired) | ||
CR10 (retired) | ||
CR10X (retired) | ||
CR200X (retired) | ||
CR206X (retired) | ||
CR211X (retired) | ||
CR216X (retired) | ||
CR23X (retired) | ||
CR295X (retired) | ||
CR300 (retired) | ||
CR3000 (retired) | ||
CR500 (retired) | ||
CR5000 (retired) | ||
CR510 (retired) | ||
CR6 | ||
CR800 (retired) | ||
CR850 (retired) | ||
CR9000 (retired) |
Data Logger Model | Data Logger Operating System (OS) or PROMs Required |
CSAT3 OS Required |
21X | 1K 6145-3, 6146-5, 10006-1 | 2.1 or higher |
21X | 2K 6148-5, 6149-7, 6070-64 | 2.1 or higher |
CR800, CR850 | All OS | 3.0 or higher |
CR10 | 2K 5954-396; 4K is a Library Special |
2.1 or higher |
CR10X | OS 1.4 or higher | 2.1 or higher |
CR1000 | All OS | 3.0 or higher |
CR23X | All OS | 2.1 or higher |
CR3000 | All OS | 3.0 or higher |
CR5000 | All OS | 3.0 or higher |
CR9000 | 2.01 or higher | 2.1 or higher |
CR9000X | All OS | 2.1 or higher |
CSAT3: 10
展开全部收起全部
The CSAT3 can be mounted on the end of a CM202, CM203, CM204, or CM206 crossarm, using the included CM250 Leveling Mounting Kit. Alternatively, the CSAT3 can mounted to a flat horizontal surface using a 3/8-16 stainless-steel bolt.
The CSAT3/3A/3H is compatible with three output signals: analog, RS-232, and Synchronous Device for Measurement (SDM). The sonic anemometer can be interfaced to any data-acquisition system that is compatible with analog measurements or RS-232 serial communications. SDM is used with Campbell Scientific data loggers.
Yes. The CSAT3 electronics contain unique calibration and sensor head geometry information. The sensor head and electronics are a matched set. (The serial number on the sensor head must match the electronics to make valid measurements.)
With long cable lengths, monitor the SDM data collected from the CSAT3. If the CSAT3 diagnostic word is 61441 or NAN, there are SDM communications errors. If the SDM bit period is increased, SDM communications should be restored. Use the SDMSpeed() instruction to change the SDM bit period. Use the following values for the listed data loggers:
The sensors used in the eddy-covariance application are not compatible with SCWin. An open-path eddy-covariance program is available for purchase as pn 18442, CRBasic Basic Eddy Covariance Program. This is the program without energy balance sensors. Also available is pn 18443, CRBasic Extended Eddy Covariance Program, with energy balance sensors. To order a custom configured program, contact the Environmental Group for assistance.
The CSAT3/3A/3AH calibration applies a correction for transducer delay over temperature. Transducer delays cause an offset in the wind speed measurement, and to a lesser extent, an offset in the speed of sound measurement.
The CSAT3/3A/3AH speed of sound easement is corrected for the effects of wind blowing normal to the sonic path.
Yes. The bubble level can be replaced in the field using pn 31962, 1-Ring Replacement Bubble Level with Mounting Bumper and Hardware.
No. The CSAT3/3A/3AH is a sensor. Time stamps are assigned to the sonic anemometer data by the data-acquisition system—typically a Campbell Scientific data logger or PC.
Sensible heat flux is defined as the covariance of vertical wind and temperature fluctuations, measured with fast-response sensors. For more information, refer to a micrometeorology textbook.
The CSAT3/3A/3AH should not be used to measure absolute sonic temperature. Aside from the effects of water vapor in the sonic path, the speed of sound measurement is highly dependent on the ability of the sonic anemometer to maintain its geometry. A 0.2 mm change in the sonic path is equal to a 1°C change in the measured absolute sonic temperature at 25°C. This is equal to a 0.33% error in the absolute sonic temperature. The wind speed measurements are also dependent on the sonic path distance, but to a lesser degree.