EE181 是一款空气温度和相对湿度RH探头,适合无人值守的应用。RH元件的专有涂层使其免受灰尘、盐或其它污染物影响,延长使用寿命。采用1000Ω的PRT铂电阻测量空气温度:-40° 至 +60°C。 EE181 建议每年校准一次。
This featured product was part of the Aconcagua expedition to install a weather station on South America’s tallest peak. Learn more about the expedition on our Extreme Application Series: Aconcagua page.
| Measurement Description | Temperature, relative humidity |
| Signal Type/Output | Analog voltage |
| Supply Voltage | 7 to 30 Vdc (typically powered by the datalogger’s 12 V supply) |
| Average Current Consumption | < 1.2 mA |
| Filter Description | 30 μm pore size, stainless-steel mesh |
| Startup Time | 2 s |
| Housing Body Material | Plastic |
| Housing Classification | IP65 |
| Field Replaceable Chip or Recalibrate | Recalibrate |
| Operating Temperature Range | -40° to +60°C |
| Sensor Diameter | 2.1 cm (0.83 in.) |
| Length | 16.0 cm (6.3 in.) |
| Weight | 290 g (10.2 oz) with 5 m cable |
Air Temperature |
|
| Sensing Element | 1000 Ω Platinum Resistance Thermometer (PRT) |
| Measurement Range | -40° to +60°C |
| Storage Temperature Range | -40° to +80°C |
| Output Signal Range | 0 to 1 Vdc |
| Accuracy | ±0.2°C (at +23°C) |
Relative Humidity |
|
| Sensing Element | Capacitance |
| Measurement Range | 0 to 100% RH (non-condensing) |
| Output Signal Range | 0 to 1 Vdc |
| Temperature Dependence | Typically 0.03% RH/°C |
| Accuracy |
|
Please note: The following shows notable compatibility information. It is not a comprehensive list of all compatible products.
| Product | Compatible | Note |
|---|---|---|
| CR1000X (retired) | ||
| CR300 (retired) | ||
| CR3000 (retired) | ||
| CR310 | ||
| CR350 | ||
| CR6 | ||
| CR800 (retired) | ||
| CR850 (retired) |
When exposed to sunlight, the EE181 must be housed in a 10-plate solar radiation shield. Campbell Scientific recommends the MetSpec RAD10E, which uses a double-louvered design that offers improved sensor protection from driving rain, snow, and insect intrusion. This shield also has lower self-heating in bright sunlight combined with higher temperatures (> 24°C [~75°F]) and low wind speeds (< 2 m s-1 [~4.5 mph]), giving a better measurement.
The EE181 will work with R. M. Young 10-plate radiation shields (41003-5 or 41003-5A), but these shields require a special adapter (pn 28415) to accommodate the sensor girth.
The RAD10E and 41003-5 attach to a crossarm, mast, or user-supplied pipe with a 2.5 to 5.3 cm (1.0 to 2.1 in.) outer diameter. The 41003-5A radiation shield attaches to a CM500-series pole or a user-supplied pole with a 5.1 cm (2.4 in.) outer diameter.