4WFBS1K 是4线全桥模块,作为单变量电阻为应变计或其它传感器提供桥臂电阻,该模块用于那些有1 kohm总阻抗的传感器。
4WFBS1K 是一个4线全桥桥臂模块。可用来组成一个全单臂电桥,亦可用来组成后半段单臂电桥。数据采集器可以测量出电桥电路为输出电压增加阻抗而产生的微小的变化。
4WFBS1K 包括2个外部的针,用于分路校正,通常被用来校正比较微弱的误差。
Used With | 4-wire strain gages or other full bridge measurements that have a 1 kΩ nominal resistance |
Dimensions | 1.5 x 1.5 x 2.7 cm (0.6 x 0.6 x 1.0 in.) for body with prongs |
Weight | 6 g (0.2 oz) |
2:1 Resistive Divider |
|
Resistance | 1 kΩ/1 kΩ |
Tolerance | Ratio: ±0.01% (@ 25°C) |
Maximum Temperature Coefficient | Ratio: 0.5 ppm/°C (-55° to +85°C) |
Power Rating | 0.1 W (@ 70°C) per element |
Completion Resistor (bulk metal foil) |
|
Resistance | 1 kΩ |
Tolerance | ±0.01% (@ 25°C) |
Maximum Temperature Coefficient | 0.8 ppm/°C |
Power Rating | 0.25 W (@ 70°C) |
Please note: The following shows notable compatibility information. It is not a comprehensive list of all compatible products.
Product | Compatible | Note |
---|---|---|
21X (retired) | ||
CR10 (retired) | ||
CR10X (retired) | ||
CR200X (retired) | ||
CR206X (retired) | ||
CR211X (retired) | ||
CR216X (retired) | ||
CR23X (retired) | ||
CR295X (retired) | ||
CR300 (retired) | ||
CR3000 (retired) | ||
CR310 | ||
CR350 | ||
CR500 (retired) | ||
CR5000 (retired) | ||
CR510 (retired) | ||
CR6 | ||
CR800 (retired) | ||
CR850 (retired) |
Product | Compatible | Note |
---|---|---|
CR9050 (retired) |
The 4WFBS1K is compatible with our multiplexers.
The 4WFBS1K uses one differential analog input; an adjacent analog ground channel accepts the "ground prong" of the 4WFBS1K. Not all data logger terminal strips have this sequence on all channels, so please check your data logger's wiring panel to confirm channel assignments, especially if you plan to use multiple Terminal Input Modules.
The lead wire that emanates from the head of the 4WFBS1K connects to a data logger excitation channel.
Please note that if a CR10X will be used to measure thermocouples in the same application, newer CR10X Wiring Panels and the CR10XTCR Reference Thermistor (with its accompanying Thermocouple Cover) are deep enough to cover the data logger input channels with a Terminal Input Module attached; older CR10TCRs are not deep enough.
The 4WFBS1K is measured with the BrFull Instruction in CRBasic, and with Instruction 6 (Full Bridge) in Edlog.
4WFBS1K: 4
展开全部收起全部
The CR1000 excitation channel current limit is 25 mA at 2500 mV. Using 1 kohm gages with multiple 4WFBS1K modules, 10 strain gages can be hooked up for each excitation channel with 2500 mV excitation.
Each data logger has specific excitation current sourcing capability. It is imperative not to exceed this current limit when trying to excite multiple strain gages from the same excitation channel.
More than one strain gage sensor can use the same excitation terminal if the bridge resistance is high enough or the excitation voltage is limited. The limit is determined by the current capacity of the excitation terminal (25 mA at 2.5 V) and the equivalent resistance of the circuit.
To make strain gage measurements, a terminal input module is needed. It is necessary to match the resistance of the strain gage with the resistance of the terminal input module.