Services Available | |
---|---|
Repair | No |
Calibration | Yes |
Free Support | No |
HMP45C 是一咱坚实耐用的精确的空气温度/相对湿度探头,它测量0 到 100% RH范围内的相对湿度,和-40° to +60°C范围内的温度。这种探头适用于长期的无人值守的监测应用,与所有的坎贝尔数据采集器都兼容。
HMP45C 使用HUMICAP电容型薄膜多聚物传感器来测量相对湿度,和一个PRT传感器来测量温度。建议每年对HMP45C重校准一次
为了减小功耗,可以仅在测量期间给HMP45C供电,此时可应用数据采集器的开关12 V端子。对于没有开关12 V端子的数据采集器,例如CR510 或 CR7,可使用SW12V 开关12 V模块来切换控制HMP45C传感器的供电。
Supply Voltage | 12 Vdc nominal (typically powered by data logger) |
Current Drain | ≤ 4 mA (active) |
Cable Diameter | 0.8 cm (0.3 in.) |
Sensor Diameter | 2.5 cm (1 in.) |
Sensor Length | 25.4 cm (10 in.) |
Weight | 0.27 kg (0.6 lb) |
Relative Humidity |
|
Sensor | Vaisala’s HUMICAP H-chip |
Measurement Range | 0.8% to 100% RH (non-condensing) |
Output Signal Range | 0.008 to 1 Vdc |
Accuracy at 20°C (against factory reference) | ±1% RH |
Accuracy at 20°C (field-calibrated against references) |
|
Temperature Dependence | ±0.05% RH/°C |
Long-Term Stability | Typically, better than 1% RH per year |
Response Time | 15 s with membrane filter (at 20°C, 90% response) |
Settling Time | 500 ms |
Temperature |
|
Temperature Sensor | 1000 ohm Platinum Resistance Thermometer |
Measurement Range | -39.2° to +60°C |
Output Signal Range | 0.008 to 1.0 V |
Accuracy at -40°C | ±0.5°C |
Accuracy at -20°C | ±0.4°C |
Accuracy at 0°C | ±0.3°C |
Accuracy at 20°C | ±0.2°C |
Accuracy at 40°C | ±0.3°C |
Accuracy at 60°C | ±0.4°C |
Please note: The following shows notable compatibility information. It is not a comprehensive list of all compatible products.
Product | Compatible | Note |
---|---|---|
21X (retired) | ||
CR10 (retired) | ||
CR10X (retired) | ||
CR200X (retired) | ||
CR206X (retired) | ||
CR211X (retired) | ||
CR216X (retired) | ||
CR23X (retired) | ||
CR295X (retired) | ||
CR3000 (retired) | ||
CR500 (retired) | ||
CR5000 (retired) | ||
CR510 (retired) | ||
CR800 (retired) | ||
CR850 (retired) | ||
CR9000 (retired) |
The HMP45C should be housed in a 41003-5 10-plate naturally-aspirated solar radiation shield. The 27251 hex plug is required to mount the HMP45C in the 41003-5. At one time, this hex plug was shipped with the radiation shield but it no longer is. The41003-5 mounts to a tripod or tower mast, crossarm, or user-supplied pole.
HMP45C-L: 1
The HMP60-L (with the Campbell Scientific cable) cannot be used with the 41003-5 10-Plate Radiation Shield without the use of additional hardware. For more information, refer to the Compatibility information on the HMP60-L product page. The standard radiation shield for the HMP60-L is the 41303-5A.