CFM100将数据采集器中的数据存储在可移除的CF卡中。CFM100/CF卡的组合可扩展数据采集器的存储能力,将野外站点的数据或程序转移到办公室,以及上载启动功能。CFM100带有40针的外设端口,可连接CR1000或CR3000数据采集器。
CFM100包括一个卡槽,可以放入I型或II型CF卡。只有工业级别的CF卡才适用于我们的产品。尽管用户级别的CF卡较便宜,但此类CF卡易受故障影响从而丢失数据或损坏卡的性能。工业级别的CF卡较用户级别的卡有更宽的温度承受范围,使用寿命也更长。
CF可取出,卡上存储的数据可通过电脑或连接数据采集器得到。使用电脑上的PCMCIA卡槽、CF1适配器、USB端口和17752读写器可读取CF卡上的数据。
卡转换程序用于从CF卡中转换和保存二进制的数据。它内置在PC200W, PC400, LoggerNet和RTDAQ数据采集器软件中。
Storage Capacity | 256 MB, 2 GB, or 16 GB |
Typical Access Speed | 200 to 400 kbits s-1 |
Memory Configuration | User selectable; ring (default) or fill-and-stop |
Power Requirements | 12 V supplied through the data logger’s peripheral port |
CF Card Requirements | Industrial-grade |
Temperature Range |
|
Dimensions |
|
Weight | 133 g (4.7 oz) |
Typical Current Drain |
|
RS-232 Port Active |
|
RS-232 Port Not Active |
|
Low Power Standby | 700 to 800 μA |
Please note: The following shows notable compatibility information. It is not a comprehensive list of all compatible products.
Product | Compatible | Note |
---|---|---|
21X (retired) | ||
CR10 (retired) | ||
CR1000X (retired) | ||
CR10X (retired) | ||
CR200X (retired) | ||
CR206X (retired) | ||
CR211X (retired) | ||
CR216X (retired) | ||
CR23X (retired) | ||
CR295X (retired) | ||
CR300 (retired) | ||
CR3000 (retired) | ||
CR310 | ||
CR500 (retired) | ||
CR5000 (retired) | ||
CR510 (retired) | ||
CR6 | ||
CR800 (retired) | ||
CR850 (retired) | ||
CR9000 (retired) |
Product | Compatible | Note |
---|---|---|
ENC10/12 | Compatible when attached to the CR1000. The CFM100 is compatible with the CR3000 using the -NB option (base with no battery). | |
ENC10/12R | Compatible when attached to the CR1000 or CR3000. | |
ENC12/14 | Compatible when attached to the CR1000 or CR3000. | |
ENC14/16 | Compatible when attached to the CR1000 or CR3000. | |
ENC16/18 | Compatible when attached to the CR1000 or CR3000. | |
ENC24/30 | Compatible when attached to the CR1000 or CR3000. | |
ENC24/30S | Compatible when attached to the CR1000 or CR3000. |
Campbell Scientific offers and recommends several CF cards (see Ordering tab). These cards have passed Campbell Scientific's ESD testing and operate properly with our data loggers.
Only industrial-grade CF card should be used with our products. Although consumer-grade cards cost less than industrial-grade cards, the consumer-grade cards are more susceptible to failure resulting in both the loss of the card and its stored data. Industrial-grade cards also function over wider temperature ranges and have longer life spans than consumer-grade cards.
Data stored on the card can be retrieved either through a communications link with the data logger or by removing the card and carrying it to a computer. The computer can read the CF card either with the computer's PCMCIA slot and the CF1 adapter or the computer's USB port and the 17752 Reader/Writer.
The CFM100 fastens onto the data logger's 40-pin peripheral port. This means that the CFM100 is only compatible with the CR1000 and CR3000. It is NOT compatible with the CR800, CR850, CR200X, CR5000, or CR9000X dataloggers.
To read cards with more than 2 GB of storage, the data logger operating system must be OS 25 or later.
CFM100: 20
展开全部收起全部
The file on the card is open until the program is stopped or the button on the device is pressed.
The CFM100 does not have any diagnostics that can be reported. The Status table does report the status of the CompactFlash card and the number of bytes free on the CF card.
Review the CRBasic Editor Help for information on the TableFile() instruction.
No. Only a program change could cause this. To save data to a CFM100 or an NL115, a CardOut() or TableFile() instruction must be present in at least one data table.
If the CardOut() or TableFile() instructions are configured as fill and stop, the data logger will stop writing data to the CompactFlash cards after they are full.
The best practice is to format the CF card, not using “Quick Format” on a PC, before inserting it into a device.
LoggerNet collects data from the data logger first. If there is additional data on a CompactFlash card, LoggerNet collects that also. Nothing special needs to be set up in LoggerNet to have this occur.
The CFM100 or NL115 is detected during the data logger boot-up cycle, that is, when it is powered on. Disconnect or connect the CFM100 or NL115 to a data logger only after power to the data logger has been disconnected.
The SC115 works with any Campbell Scientific data logger that has a CS I/O terminal.